Skip to main content
U.S. flag

An official website of the United States government

Evaluation on determination of iodine in coal by energy dispersive X-ray fluorescence

January 1, 2005

A quick and inexpensive method of relative high iodine determination from coal samples was evaluated. Energy dispersive X-ray fluorescence (EDXRF) provided a detection limit of about 14 ppm (3 times of standard deviations of the blank sample), without any complex sample preparation. An analytical relative standard deviation of 16% was readily attainable for coal samples. Under optimum conditions, coal samples with iodine concentrations higher than 5 ppm can be determined using this EDXRF method. For the time being, due to the general iodine concentrations of coal samples lower than 5 ppm, except for some high iodine content coal, this method can not effectively been used for iodine determination. More work needed to meet the requirement of determination of iodine from coal samples for this method. Copyright ?? 2005 by The Geochemical Society of Japan.

Publication Year 2005
Title Evaluation on determination of iodine in coal by energy dispersive X-ray fluorescence
DOI 10.2343/geochemj.39.391
Authors B. Wang, J.C. Jackson, C. Palmer, B. Zheng, R. B. Finkelman
Publication Type Article
Publication Subtype Journal Article
Series Title Geochemical Journal
Index ID 70029606
Record Source USGS Publications Warehouse
Was this page helpful?