Multi-saline sample distillation apparatus for hydrogen isotope analyses: Design and accuracy
A distillation apparatus for saline water samples was designed and tested. Six samples may be distilled simultaneously. The temperature was maintained at 400 C to ensure complete dehydration of the precipitating salts. Consequently, the error in the measured ratio of stable hydrogen isotopes resulting from incomplete dehydration of hydrated salts during distillation was eliminated. (USGS)
Citation Information
Publication Year | 1981 |
---|---|
Title | Multi-saline sample distillation apparatus for hydrogen isotope analyses: Design and accuracy |
DOI | 10.3133/wri8125 |
Authors | Afifa Afifi Hassan |
Publication Type | Report |
Publication Subtype | USGS Numbered Series |
Series Title | Water-Resources Investigations Report |
Series Number | 81-25 |
Index ID | wri8125 |
Record Source | USGS Publications Warehouse |