Rapid and precise determination of ATP using a modified photometer
An inexpensive delay timer was designed to modify a commercially available ATP photometer which allows a disposable tip pipette to be used for injecting either enzyme or sample into the reaction cuvette. The disposable tip pipette is as precise and accurate as a fixed-needle syringe but eliminates the problem of sample contamination and decreases analytical time. (USGS)
Citation Information
Publication Year | 1980 |
---|---|
Title | Rapid and precise determination of ATP using a modified photometer |
DOI | 10.3133/ofr801194 |
Authors | David J. Shultz, Doyle W. Stephens |
Publication Type | Report |
Publication Subtype | USGS Numbered Series |
Series Title | Open-File Report |
Series Number | 80-1194 |
Index ID | ofr801194 |
Record Source | USGS Publications Warehouse |